-86 Prime Series Ultra Low Temperature Freezer
Operating Temperature -86 , Range -20 to -86 Air-cool refrigeration system are engineered to maintain design temperature in warm 32 ambient conditions Refrigeration System Stable cascade refrigeration system with two 1HP air-cooled compressors. Low temperature stage uses R508B, CFC-free refrigerants. Extra large condenser with two aerodynamically axial fans provides increase heat rejection. Down flow evaporator for most efficient refrigerant flow, oil return, and longer compressor life. Washable condenser filter keeps fins free of dust to maintain optimal cooling efficiency; no tools required for removal Control System Microprocessor control system with touch pad data entry and digital display of all function. Password security system provides best bulwark for main power, temperature & alarm setting. Independent operating temperature and high/low limit alarm. Factory setting for quickly setup and start. Temperature increase/decrease adjust 1 . Electrical System Automatic voltage compensates for high and low voltage situations. Online incoming line voltage monitoring and digital read-out. Built-in surge suppressor minimizes effect of voltage spikes from incoming electrical service. solid state relay driver, no moving component in compressor driver circuit, no spike in start situation to protect microprocessor control. Build in current breaker, protect compressor and power supply circuit. Build in Automatic Delay Restart (ADR), ensure the on line quality after power failure, decrease power circle loading, and protect compressor.